Sains Malaysiana 38(2):
197-202(2009)
Absorbed Dose to Water Intercomparison Programme
for Radiotherapy Centres
(Program Antarabandingan Dos
Terserap Kepada Air
Untuk Pusat Radioterapi)
S. B. Samat & S. X. Too
School of Applied Physics, Universiti Kebangsaan
Malaysia (UKM)
43600 UKM Bangi, Selangor
T. Kadni & M.T. Dolah
Secondary Standard Dosimetry Laboratory (SSDL)
Malaysian Nuclear Agency (Nuclear Malaysia)
43000 Kajang, Selangor
Received: 8 May 2008 / Accepted: 20 June 2008
ABSTRACT
The
importance of intercomparison programme for radiation therapy dose delivery at
radiotherapy centres (RC) has widely been reported. We at Nuclear Malaysia
Agency (NM) intend to develop a national level intercomparison programme in
terms of absorbed dose to water Dw. We started with one RC which has
voluntered to participate in the programme in the dose delivery of 6 MV x-ray
beam. Two types of comparison were made to get the accuracy. Type I is based on
the measured value of the Dw at reference depth, Dw(zref)mea : RC results was compared with NM, in which NM results was taken as a
standard. Type II is based on calculated value the Dw(zref),
Dw(zref)cal : both NM and RC results were
compared with Dw(zref)cal, in which Dw(zref)cal was taken as a standard. Two on-site measurements were made at this centre to
check for results consistency and to get the average accuracy. Results obtained
were consistent for the two measurements and the average values in the accuracy
are within the acceptable accuracy limit of ± 3%, namely Type I: 1.10 ±
0.58 % for the RC, and Type II: 0.99 ± 0.60 % and −0.11 ± 0.26%
respectively for the RC and NM.
Keywords:
Absorbed dose to water; intercomparison
programme; radiotherapy centres
ABSTRAK
Kepentingan program antarabandingan untuk
pemberian dos sinaran terapi di pusat radioterapi (RC) telah banyak dilaporkan. Kami di Agensi Nuklear Malaysia (NM)
berhasrat untuk membangunkan program antarabandingan peringkat kebangsaan dalam
sebutan dos terserap kepada air Dw. Kami bermula dengan satu RC yang
secara sukarela hendak menyertai program ini dalam pemberian dos alur sinar-x 6
MV. Dua jenis perbandingan dibuat untuk
mendapatkan ketepatan. Jenis I berdasarkan nilai pengukuran Dw pada
kedalaman rujukan, Dw(zref)mea: keputusan RC dibandingkan
dengan NM, dengan keputusan NM diambil sebagai standard. Jenis II berdasarkan
nilai kiraan Dw(zref), Dw(zref)cal : kedua-dua keputusan NM dan RC dibandingkan dengan Dw(zref)cal,
dengan mengambil Dw(zref)cal sebagai standard. Dua
pengukuran di tempat kajian dilakukan dipusat ini bagi tujuan penyemakan
kekonsistenan dan mendapatkan purata ketepatan. Keputusan didapati adalah
konsisten bagi dua pengukuran itu dan purata ketepatan berada di dalam had
ketepatan yang diterima ± 3%, iaitu Jenis I: 1.10 ± 0.58 % untuk RC, dan Jenis
II: 0.99 ± 0.60 % dan −0.11 ± 0.26% masing-masing untuk RC dan NM.
Kata kunci: Dos terserap kepada air; program antarabandingan; pusat radioterapi
REFERENCES
Amat, A.
2006. Perbandingan Ketepatan dan Kepersisan Nilai Dos Terserap Kepada Air
Ditentukan Dengan Protokol IAEA TRS-398 Dab AAPM TG-51 Pada Alur Sinar-X 6 MV. Tesis
Sarjana Sains, Pusat Pengajian Fizik Gunaan, Universiti Kebangsaan Malaysia.
BJR. 1983. Central Axis Depth Dose Data For Use In Radiotherapy: A Survey Depth
Dose And Related Data Measured In Water Or Equivalent Media, Supplement No: 17. London: The British Institute Of Radiology.
IAEA. 1997a. Absorbed Dose Determination In Photons And Electron Beams:
An International Code Of Practice, TRS-277. Ed. 2. Vienna: International
Atomic Energy Agency.
IAEA. 1997b. IAEA Certificate For Secondary Standard Ionization Chamber, Ref. No.
MAL/INTER/97. Vienna: International Atomic Energy Agency.
IAEA. 2000. Absorbed Dose Determination In External Beam Radiotherapy: An
International Code Of Practice For Dosimetry Based On Standards Of Absorbed
Dose To Water, TRS-398. Vienna: International Atomic Energy Agency.
Izewska, J., Svensson, H.
& Ibbott, G. 2002. Worldwide QA Networks
For Radiotherapy Dosimetry. Proceedings Of
An International Symposium On Standards And Codes Of Practice In Medical
Radiation Dosimetry. Vol. 2: 139-155.
Samat, S.B., Too, S.X., Kadni, T.
& Dolah, M.T. 2007. The Performance
Stability Of ND,W Calibration
Factor For Chambers Belonging To Local Radiotherapy Centres, Proceedings Of Malaysian Science And Technology Congress 2007 (MSTC 2007):172-177.
Too, S.X., Samat, S.B., Kadni, T.
& Dolah, M.T. 2007. Procedures For Dw TRS-398 Intercomparison
Programme Between Malaysian Radiotherapy Centres And SSDL Malaysia, Presented At Persidangan Fizik Kebangsaan 2007 (PERFIK 2007), Kuala Terengganu.
Wikipedia. 2008. Monitor Unit. Http://En.Wikipedia.Org/Wiki/Monitor_Unit. [28 April 2008].
|