Sains Malaysiana 41(1)(2012): 133–140

 

EDR2 Film for Skin Dose Measurement in Coronary Angiography

(Pengukuran Dos Kulit dalam Koronari Angiografi Menggunakan Filem EDR2)

 

 

C.K. Ying & S. Kandaiya*

School of Physics, Universiti Sains Malaysia, 11700, Gelugor, Penang, Malaysia

 

Received: 2 March 2011 / Accepted: 10 June 2011

 

 

ABSTRACT

 

Patient skin dose measurements were performed at the cardiology department in a hospital in Penang, Malaysia using EDR films during coronary angiography (CA) procedures. The EDR2 film was first characterized in terms of dose and energy dependence as well reproducibility. For patient dose measurements, the films were placed on the table underneath the patient for an under couch tube position. A total of 27 CA procedures were studied. Results for peak skin doses (PSD) ranged from 35- 684 mGy while the dose area product (DAP) values were from 5.5 to 93.1 Gy cm2. DAP correlated reasonably with PSD for CA procedure (R2= 0.8212). The highest PSD value in this study is below the threshold dose value of 2 Gy for early transient skin injury recommended by the Food and Drug Administration (FDA 1994).

 

Keywords: Area Product; dose peak skin dose; EDR2 Film; interventional Cardiology

 

ABSTRAK

Pengukuran sinaran dos pesakit telah dijalankan di bahagian kardiologi, sebuah hospital di Pulau Pinang, Malaysia dengan menggunakan filem EDR2 semasa menjalankan prosedur koronari angiografi(CA). Pencirian filem EDR2 pada mulanya dilakukan dalam sebutan kebergantungan dos dan tenaga dan juga keupayaan kebolehpenghasilan semula. Untuk mengukur dos pesakit, filem EDR2 diletakkan di atas meja dan di bawah badan pesakit dengan tiub X-ray yang terletak di bawah meja. Sebanyak 27 prosedur CA dikaji. Keputusan nilai puncak dos kulit (PSD) adalah dalam julat 35 – 684 mGy, sementara pendaraban dos dengan luas (DAP) pula adalah dalam julat 5.5 – 93.1 Gy cm2. DAP berkorelasi secara wajar dengan PSD dalam prosedur CA (R2 = 0.8212). Nilai PSD maxima di dalam penyelidikan ini adalah di bawah nilai ambang dos 2 Gy untuk kecederaan kulit yang dicadangkan oleh Pentadbiran Makanan dan Ubat-ubatan (FDA 1994).

 

Kata kunci: Filem EDR2; intervensi kardiologi; pendaraban dos dengan luas; puncak dos kulit

 

REFERENCES

 

Balter, S. 2006. Method for measuring fluoroscopic skin dose. Pediatric Radiology 36: 136-140.

Betsou, S., Efstathopoulos, E.P., Katritsis, D., Faulkner, K. & Panayiotakis, G. 1998. Patient radiation doses during cardiac catheterization procedures. British Journal of Radiology 71: 634-39.

Canne, S.D., Carosi, A., Bufacchi, A., Malatesta, T., Capparella, R., Fragomeni, R., Adoranre, N., Bianchi, S. & Begnozzi, L. 2006. Use of GAFCHROMIC XR Type R films for skin-dose measurement in interventional radiology: Valodation of a dosimetric procedure on a sample of patients undergone interventional cardiology. Physica Medica 22: 105-110.

Delichas, M.G., Psarrakos, K., Hatziioannou, K., Giannoglou, G., Molyvda-Athanasopoulou, E., Papanastassiou, E. & Sioundas, A. 2005. The dependence of patient dose on factors relating to the technique and complexity of international cardiology procedure. Physica Medica 21: 153-157.

Dogan, B., Olğar, T., Toklu, T., Cağlan, A., Onal, E. & Padovani, R. 2008. Patient doses and dosimetric evaluations in interventional cardiology. Physica Medica 20: 1-12.

FDA US Food & Drug Administration 1994. Avoidance of serious x-ray induced skin injuries to patient during fluoroscopically-guided procedures. Medical Bulletin 24(2): 7-17.

Fletcer, D.W., Miller, D.L., Balter, S. & Taylor, M.S. 2002. Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures. Journal of Vascular and Interventional Radiology 13: 391-7.

Guibelalde, E., Vano, E., Gonzalez, L., Prieto, C., Fernandez, J.M. & Ten, J.I. 2003. Practical aspects for the evaluation of skin doses in interventional cardiology using a new slow film. British Journal of Radiology 76: 332-36.

Hart, D. & Wall, B.F. 2002. Radiation Exposure of the UK Population from Medical and Dental X-ray Examinations Chilton. UK NRPB-W4.

Khan, Faiz. M. 2003. The physics of Radiation Therapy. (3rd ed). Baltimore, USA: Williams & Wilkins.

Le-Heron, J.C. 1992. Estimation of effective dose to the patient during the medical X-ray examinations from measurement of the dose-area product. Physics in Medicine and Biology 37: 2117-26.

McParland, B.J. 1998. Entrance skin dose estimates derived from dose-area product measurements in interventional radiological procedures. British Journal of Radiology 71: 1288-95.

Morrell, R.E. & Rogers, A. 2004. Calibration of Kodak EDR2 film for patient skin dose assessment in cardiac catheterization procedures. Physics in Medicine and Biology 49: 5559-70.

Morrell, R.E. & Rogers, A. 2006. Kodak EDR2 film for patient skin dose assessment in cardiac catheterization procedures. British Journal of Radiology 79: 603-7.

Morrish, O.W.E. & Goldstone, K.E. 2008. An investigation into patient and staff doses from x-ray angiography during coronary interventional procedures. British Journal of Radiology 81: 35-45.

Ng, K.H., Abdullah, B.J. & Sivalingam, S. 1999. Medical radiation exposures for diagnostic radiology in Malaysia. Health Physics 77: 33-6.

Padovani, R., Novario, R. & Bernardi, G. 1998. Optimization in coronary angiography and percutaneous transluminal coronary angioplasty. Radiation Protection Dosimetry 80: 303-6.

Sapiin, B., Ng, K.H. & Abdullah, B.J.J. 2004. Radiation dose to patients undergoing interventional radiological procedures in selected hospitals in Malaysia: Retrospective Study. Journal of the Hong Kong College of Radiology 7: 129-136.

Tsapaki, V., Kottou, S., Vano, E., Faulkner, K., Giannouleas, J., Padovani, R., Kyrozi, E., Koutelou, M., Vardalaki, E. & Neofotistou, V. 2003. Patient dose values in a dedicated Greek cardiac centre. British Journal of Radiology 76: 726-30.

Van de Putte, S., Verhaegen, F., Taeymans, Y. & Thierens, H. 2000. Correlation of patient skin doses in cardiac interventional radiology with dose-area product. British Journal of Radiology 73: 504-13.

Vano, E., González, L., Fernández, J.M. & Guibelalde, E. 1995. Patient doses in interventional radiology. British Journal of Radiology 68: 1215-20.

Vano, E., Guibelalde, E., Fernandez, J.M., Gonzalez, L. & Ten, J.I. 1997. Patient dosimetry in interventional radiology using slow film. British Journal of Radiology 70: 195-200.

 

*Corresponding author; email: sivamany@usm.my

 

 

 

 

 

 

 

 

 

previous